![]() ![]() A fresh 21 Gauze Olympus ViziShot Needle (model NA-201SX-4021) was introduced through the working channel of the EBUS bronchoscope. The lymph node was clearly visualized which had a homogenous echogenicity. The ultrasound transducer was directly coupled with the bronchial wall without inflating the saline balloon to target the lymph node at station 7 just distal to carina in the right main bronchus. Linear EBUS bronchoscopy was performed using a laryngeal mask airway under general anesthesia. A contrast-enhanced computed tomography of the chest revealed a homogenous subcarinal lymph node measuring 20 mm × 25 mm with contrast enhancement. Here, we describe a rare complication of needle assembly malfunction due to the separation of the shaft of needle and sheath-sliding mechanism which resulted in the failure to retract the needle within the sheath during the procedure.Ī 43-year-old male presented with a history of fever and cough for 3 weeks. The endobronchial ultrasound (EBUS) scope is delicate equipment through which a dedicated needle assembly is passed for performing the transbronchial needle aspiration.